Question 10 of 20 Fundamentals Practical
How do you titrate a flow cytometry antibody and pick the concentration to use?
Short answer
Stain cells with a serial dilution of the antibody under your real staining conditions, then choose the concentration that best separates positive from negative cells. Too much antibody raises background on negative cells, and too little loses dim positives.
Why
The vendor's recommended amount was set on someone else's cells, buffer, time and temperature. The Cossarizza guidelines therefore recommend titrating every reagent, and every new lot, in the conditions of the actual experiment. Separation is measured on both populations.
The stain index compares the gap between positive and negative medians with the spread of the negatives. It falls at high concentrations because the negatives brighten, and at low ones because the positives dim. A population that lacks the antigen is needed in the sample to read background.
Concentration, not total amount, is the key variable, so scaling up the staining volume needs proportionally more antibody. Staining is robust within a fairly wide range of cell numbers. Fixed and live cells, or surface and intracellular staining, can have different optima.
How to work it through
- Use the same cell type, cell number, buffer, volume, time and temperature as the real assay.
- Start from the vendor's concentration or a generic high point and make six to eight twofold dilutions.
- Include unstained cells and make sure a negative population is present.
- Acquire all tubes with identical instrument settings.
- Plot the median of positives, the median of negatives and the stain index against concentration.
- Choose the concentration with the best separation, or the lowest one that reaches near-saturating signal if robustness matters more.
An example
A CD8 antibody is tested from 4 down to 0.06 micrograms per test. At 4 micrograms the negative median is 900 and the positive 52,000. At 0.5 micrograms the negative median falls to 220 while the positive is still 48,000, so the stain index is highest there.
Below 0.25 micrograms the positives drop sharply. The lab adopts 0.5 micrograms, an eightfold saving, with better separation than the vendor's dose.
Where the answer stops holding
An optimum found in a single-color tube can shift inside a full panel, where spread from other dyes changes the negative population. Titration on a cell line may not hold for primary tissue with different antigen density. Very rare targets may need saturating staining for consistency across samples, even at some cost in background.
A strong answer covers
- Titrate in real staining conditions and per lot
- Separation, not brightness of positives, is the criterion
- Stain index captures both background and signal
- Concentration rather than amount is what matters
- Re-check the optimum inside the full panel
Common mistakes
- Using the vendor's recommended dose without titrating
- Picking the concentration that gives the brightest positive population regardless of background
- Titrating on cells that contain no negative population to measure background
The interviewer might ask next
Why might the best concentration for a surface stain differ from the best concentration after fixation?