How do primary and secondary reference standards differ, and how is each qualified?
Short answer
A primary reference standard comes from an official source such as a pharmacopoeia, or is fully characterized in house when none exists. A secondary or working standard is qualified against the primary before first use and requalified periodically. Every result depends on the standard's assigned value.
Why
Most QC results are relative. An HPLC assay compares the sample's peak with a standard's peak, so an error in the standard's assigned purity moves every result by the same amount, and no amount of method precision will reveal it. ICH Q7 sets out the hierarchy.
Primary standards from an officially recognized source are normally used without testing if stored as the supplier recommends, with their source, storage and use recorded. Where no official standard exists, an in-house primary standard must be established with testing that fully establishes identity and purity. Secondary standards are prepared, identified, tested, approved and stored under control, compared with the primary before first use, and periodically requalified under a written protocol.
US regulation also requires complete records of testing and standardization of reference standards. Biologics add a twist, because potency is often defined by the standard itself, so replacing a bioassay reference standard needs a bridging study to avoid a step change in reported potency. For example, if a new working standard is assigned 99.2 percent when it is really 98.0 percent, every assay result reads about 1.2 percent low.
The limit of this system is that it relies on careful storage and handling. A hygroscopic standard left open absorbs water and quietly changes its effective purity.
A strong answer covers
- Official primary standards used as supplied when stored correctly
- In-house primary needs full identity and purity characterization
- Secondary qualified against primary and periodically requalified
- Assigned value error biases every result
- Bridging when replacing biological reference standards
Common mistakes
- Assuming method precision would reveal an error in a standard's assigned value.
- Using a working standard indefinitely without requalification.
- Replacing a bioassay reference standard without a bridging study.
The interviewer might ask next
How would you investigate a sudden shift in assay results that coincides with a new working standard lot?